HIGH INTENSITY FOCUSED ULTRASOUND SYSTEM

Product ID:

Position:

Sound:

Types of:

Focal field:

Efficiency:

Installation Environment:

PRO2008

Top-mounted(Dry)

≥1000w/cm²

Annular self - focusing transducer

1.1*1.1*3.3mm³(Ellipsoid)

20 minutes -45 minutes to complete a single treatment

Room size:≥25㎡,Unilateral≥4m,High≥3m
Power Requirements:Single-phase three-wire,Standard ground,Power≥2KW
Water supply requirements:flow≥1m³/h
Water pressure:0.1-0.5Mpa

Component

Power Drive Module:
Convert the mains into high-frequency electronic power.

Treament Module:
Convert the high-frequency electronic power into high intensity ultrasound, and transmit this ultrasound for focusing on the target.

Positioning Mechanism:
Control the position of the bed and angle of the C arm, to achieve the most optimized ultrasound pathway.

Ultrasonography Monitoring Module:
Real-time guiding and monitor the treatment procedure.

Treatment Controlling Module:
Make the treatment plan and perform the procedure.

Remote Monitoring System:Through the network can be real-time monitoring equipment running.

Clinical Data Sharing Platform:
Automatically deliver the failure warning information through internet.

Medium Treatment Device:
Provide the medium water for coupling.

Security Module:
Ensure safety operation.

Failure Warning Module:
Automatically deliver the failure warning information through internet.

Ultra-wide 2D imaging and color wide view imaging scanning

Image identification and warning function

The conformal function precisely demarcates the border of the target

Real-time monitoring,images storage